Vulnerabilities
Vulnerable Software
Security Vulnerabilities - CVEs Published In February 2018
Huawei AR3200 with software V200R006C10, V200R006C11, V200R007C00, V200R007C01, V200R007C02, V200R008C00, V200R008C10, V200R008C20, V200R008C30 has an integer overflow vulnerability. The software does not sufficiently validate certain field in SCTP messages, a remote unauthenticated attacker could send a crafted SCTP message to the device. Successful exploit could system reboot.
CVSS Score
7.5
EPSS Score
0.003
Published
2018-02-15
Huawei AR3200 with software V200R006C10, V200R006C11, V200R007C00, V200R007C01, V200R007C02, V200R008C00, V200R008C10, V200R008C20, V200R008C30 has an integer overflow vulnerability. The software does not sufficiently validate certain field in SCTP messages, a remote unauthenticated attacker could send a crafted SCTP message to the device. Successful exploit could cause system reboot.
CVSS Score
7.5
EPSS Score
0.003
Published
2018-02-15
The download commit resource in Atlassian Bitbucket Server from version 5.1.0 before version 5.1.7, from version 5.2.0 before version 5.2.5, from version 5.3.0 before version 5.3.3 and from version 5.4.0 before version 5.4.1 allows remote attackers to write files to disk potentially allowing them to gain code execution, exploit CVE-2017-1000117 if a vulnerable version of git is in use, and or determine if an internal service exists via an argument injection vulnerability in the at parameter.
CVSS Score
7.5
EPSS Score
0.019
Published
2018-02-15
Various plugin servlet resources in Atlassian Bitbucket Server before version 5.3.7 (the fixed version for 5.3.x), from version 5.4.0 before 5.4.6 (the fixed version for 5.4.x), from version 5.5.0 before 5.5.6 (the fixed version for 5.5.x), from version 5.6.0 before 5.6.3 (the fixed version for 5.6.x), from version 5.7.0 before 5.7.1 (the fixed version for 5.7.x) and before 5.8.0 allow remote attackers to conduct clickjacking attacks via framing various resources that lacked clickjacking protection.
CVSS Score
4.3
EPSS Score
0.002
Published
2018-02-15
A Classic Buffer Overflow issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump does not verify input buffer size prior to copying, leading to a buffer overflow, allowing remote code execution on the target device. The pump receives the potentially malicious input infrequently and under certain conditions, increasing the difficulty of exploitation.
CVSS Score
8.1
EPSS Score
0.267
Published
2018-02-15
An Improper Access Control issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The FTP server on the pump does not require authentication if the pump is configured to allow FTP connections.
CVSS Score
8.1
EPSS Score
0.024
Published
2018-02-15
An Improper Certificate Validation issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump does not validate host certificates, leaving the pump vulnerable to a man-in-the-middle (MITM) attack.
CVSS Score
5.9
EPSS Score
0.001
Published
2018-02-15
An Out-of-bounds Read issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump reads memory out of bounds, causing the communications module to crash. Smiths Medical assesses that the crash of the communications module would not impact the operation of the therapeutic module.
CVSS Score
5.3
EPSS Score
0.008
Published
2018-02-15
A Password in Configuration File issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump stores some passwords in the configuration file, which are accessible if the pump is configured to allow external communications.
CVSS Score
3.7
EPSS Score
0.002
Published
2018-02-15
A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The FTP server on the pump contains hardcoded credentials, which are not fully initialized. The FTP server is only accessible if the pump is configured to allow FTP connections.