Vulnerability Details CVE-2017-12718

A Classic Buffer Overflow issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump does not verify input buffer size prior to copying, leading to a buffer overflow, allowing remote code execution on the target device. The pump receives the potentially malicious input infrequently and under certain conditions, increasing the difficulty of exploitation.
Exploit prediction scoring system (EPSS) score
EPSS Score 0.267
EPSS Ranking 96.1%
CVSS Severity
CVSS v3 Score 8.1
CVSS v2 Score 6.8
Products affected by CVE-2017-12718