In IntelliVue patient monitors MX100, MX400-550, MX600, MX700, MX750,
MX800, MX850, MP2-MP90, and IntelliVue X2 and X3 Versions N and prior,
the product receives input or data but does not validate or incorrectly
validates that the input has the properties required to process the data
safely and correctly, which can induce a denial-of-service condition
through a system restart.
In Patient Information Center iX (PICiX) Versions C.02, C.03,
PerformanceBridge Focal Point Version A.01, the product receives input
that is expected to be well-formed (i.e., to comply with a certain
syntax) but it does not validate or incorrectly validates that the input
complies with the syntax, causing the certificate enrollment service to
crash. It does not impact monitoring but prevents new devices from
enrolling.
In Patient Information Center iX (PICiX) Version B.02, C.02, C.03, and
PerformanceBridge Focal Point Version A.01, when an actor claims to have
a given identity, the software does not prove or insufficiently proves
the claim is correct.
In Patient Information Center iX (PICiX) Versions C.02 and C.03,
PerformanceBridge Focal Point Version A.01, IntelliVue patient monitors
MX100, MX400-MX550, MX750, MX850, and IntelliVue X3 Versions N and
prior, the software does not check or incorrectly checks the revocation
status of a certificate, which may cause it to use a compromised
certificate.