A session fixation vulnerability in the B. Braun Melsungen AG SpaceCom administrative interface Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows remote attackers to hijack web sessions and escalate privileges.
An open redirect vulnerability in the administrative interface of the B. Braun Melsungen AG SpaceCom device Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows attackers to redirect users to malicious websites.
An improper sanitization of input vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to gain user-level command-line access by passing a raw external string straight through to printf statements. The attacker is required to be on the same network as the device.
A Cleartext Transmission of Sensitive Information vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote attacker to obtain sensitive information by snooping on the network traffic. The exposed data includes critical values for a pump's internal configuration.
An Unrestricted Upload of File with Dangerous Type vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows remote attackers to upload any files to the /tmp directory of the device through the webpage API. This can result in critical files being overwritten.
An Insufficient Verification of Data Authenticity vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to send the device malicious data that will be used in place of the correct data. This results in full system command access and execution because of the lack of cryptographic signatures on critical data sets.
A Missing Authentication for Critical Function vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote attacker to reconfigure the device from an unknown source because of lack of authentication on proprietary networking commands.
A relative path traversal attack in the B. Braun OnlineSuite Version AP 3.0 and earlier allows unauthenticated attackers to upload or download arbitrary files.
A DLL hijacking vulnerability in the B. Braun OnlineSuite Version AP 3.0 and earlier allows local attackers to execute code on the system as a high privileged user.
An Excel Macro Injection vulnerability exists in the export feature in the B. Braun OnlineSuite Version AP 3.0 and earlier via multiple input fields that are mishandled in an Excel export.