Vulnerabilities
Vulnerable Software
Medtronic:  Security Vulnerabilities
Deserialization of untrusted data in Microsoft Messaging Queuing Service in Medtronic's Paceart Optima versions 1.11 and earlier on Windows allows an unauthorized user to impact a healthcare delivery organization’s Paceart Optima system cardiac device causing data to be deleted, stolen, or modified, or the Paceart Optima system being used for further network penetration via network connectivity.
CVSS Score
9.8
EPSS Score
0.297
Published
2023-06-29
Medtronic identified that the Pelvic Health clinician apps, which are installed on the Smart Programmer mobile device, have a password vulnerability that requires a security update to fix. Not updating could potentially result in unauthorized control of the clinician therapy application, which has greater control over therapy parameters than the patient app. Changes still cannot be made outside of the established therapy parameters of the programmer. For unauthorized access to occur, an individual would need physical access to the Smart Programmer.
CVSS Score
6.4
EPSS Score
0.001
Published
2023-03-01
A vulnerability exists which could allow an unauthorized user to learn aspects of the communication protocol used to pair system components while the pump is being paired with other system components. Exploitation requires nearby wireless signal proximity with the patient and the device; advanced technical knowledge is required for exploitation. Please refer to the Medtronic Product Security Bulletin for guidance
CVSS Score
4.8
EPSS Score
0.001
Published
2022-12-12
Medtronic MyCareLink Smart 25000 contains an authentication protocol vulnerability where the method used to authenticate between the MCL Smart Patient Reader and the Medtronic MyCareLink Smart mobile app is vulnerable to bypass. This vulnerability enables an attacker to use another mobile device or malicious application on the patient’s smartphone to authenticate to the patient’s Medtronic Smart Reader, fooling the device into believing it is communicating with the original Medtronic smart phone application when executed within range of Bluetooth communication.
CVSS Score
8.0
EPSS Score
0.001
Published
2020-12-14
Medtronic MyCareLink Smart 25000 is  vulnerable when an authenticated attacker runs a debug command, which can be sent to the patient reader and cause a heap overflow event within the MCL Smart Patient Reader software stack. The heap overflow could allow an attacker to remotely execute code on the MCL Smart Patient Reader, potentially leading to control of the device
CVSS Score
8.8
EPSS Score
0.009
Published
2020-12-14
Medtronic MyCareLink Smart 25000 is vulnerable to a race condition in the MCL Smart Patient Reader software update system, which allows unsigned firmware to be uploaded and executed on the Patient Reader. If exploited, an attacker could remotely execute code on the MCL Smart Patient Reader device, leading to control of the device.
CVSS Score
8.8
EPSS Score
0.003
Published
2020-12-14
In Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) version 2.1.0 and lower and version 2.0.3 and lower, and Valleylab LS10 Energy Platform (VLLS10GEN—not available in the United States) version 1.20.2 and lower, the RFID security mechanism used for authentication between the FT10/LS10 Energy Platform and instruments can be bypassed, allowing for inauthentic instruments to connect to the generator.
CVSS Score
4.8
EPSS Score
0.001
Published
2019-11-08
In Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) version 2.1.0 and lower and version 2.0.3 and lower, and Valleylab LS10 Energy Platform (VLLS10GEN—not available in the United States) version 1.20.2 and lower, the RFID security mechanism does not apply read protection, allowing for full read access of the RFID security mechanism data.
CVSS Score
4.6
EPSS Score
0.004
Published
2019-11-08
Medtronic Valleylab Exchange Client version 3.4 and below, Valleylab FT10 Energy Platform (VLFT10GEN) software version 4.0.0 and below, and Valleylab FX8 Energy Platform (VLFX8GEN) software version 1.1.0 and below use the descrypt algorithm for OS password hashing. While interactive, network-based logons are disabled, and attackers can use the other vulnerabilities within this report to obtain local shell access and access these hashes.
CVSS Score
7.0
EPSS Score
0.002
Published
2019-11-08
Medtronic Valleylab Exchange Client version 3.4 and below, Valleylab FT10 Energy Platform (VLFT10GEN) software version 4.0.0 and below, and Valleylab FX8 Energy Platform (VLFX8GEN) software version 1.1.0 and below use multiple sets of hard-coded credentials. If discovered, they can be used to read files on the device.
CVSS Score
5.8
EPSS Score
0.007
Published
2019-11-08


Contact Us