An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks
Active debug code in the B. Braun Melsungen AG SpaceCom Version L8/U61, and the Data module compactplus Versions A10 and A11 and earlier enables attackers in possession of cryptographic material to access the device as root.
A reflected cross-site scripting (XSS) vulnerability in the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows remote attackers to inject arbitrary web script or HTML into various locations.
Improper access controls in the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 enables attackers to extract and tamper with the devices network configuration.
A XPath injection vulnerability in the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows unauthenticated remote attackers to access sensitive information and escalate privileges.
A vulnerability in the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows attackers to recover user credentials of the administrative interface.
An improper verification of the cryptographic signature of firmware updates of the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows attackers to generate valid firmware updates with arbitrary content that can be used to tamper with devices.
Hard-coded credentials in the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 enable attackers with command line access to access the device’s Wi-Fi module.
A vulnerability in the configuration import mechanism of the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows attackers with command line access to the underlying Linux system to escalate privileges to the root user.
A relative path traversal attack in the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows attackers with service user privileges to upload arbitrary files. By uploading a specially crafted tar file an attacker can execute arbitrary commands.