An issue was discovered in Smiths-Medical CADD-Solis Medication Safety Software, Version 1.0; 2.0; 3.0; and 3.1. CADD-Solis Medication Safety Software grants an authenticated user elevated privileges on the SQL database, which would allow an authenticated user to modify drug libraries, add and delete users, and change user permissions. According to Smiths-Medical, physical access to the pump is required to install drug library updates.
An issue was discovered in Smiths-Medical CADD-Solis Medication Safety Software, Version 1.0; 2.0; 3.0; and 3.1. The affected software does not verify the identities at communication endpoints, which may allow a man-in-the-middle attacker to gain access to the communication channel between endpoints.
An issue was discovered in Becton, Dickinson and Company (BD) Alaris 8015 Point of Care (PC) unit, Version 9.5 and prior versions, and Version 9.7, and 8000 PC unit. An unauthorized user with physical access to an affected Alaris PC unit may be able to obtain unencrypted wireless network authentication credentials and other sensitive technical data by disassembling the PC unit and accessing the device's flash memory. The Alaris 8015 PC unit, Version 9.7, and the 8000 PC unit store wireless network authentication credentials and other sensitive technical data on internal flash memory. Accessing the internal flash memory of the affected device would require special tools to extract data and carrying out this attack at a healthcare facility would increase the likelihood of detection.
An issue was discovered in Becton, Dickinson and Company (BD) Alaris 8015 Point of Care (PC) unit, Version 9.5 and prior versions, and Version 9.7. An unauthorized user with physical access to an Alaris 8015 PC unit may be able to obtain unencrypted wireless network authentication credentials and other sensitive technical data by disassembling an Alaris 8015 PC unit and accessing the device's flash memory. Older software versions of the Alaris 8015 PC unit, Version 9.5 and prior versions, store wireless network authentication credentials and other sensitive technical data on the affected device's removable flash memory. Being able to remove the flash memory from the affected device reduces the risk of detection, allowing an attacker to extract stored data at the attacker's convenience.
IBM WebSphere Application Server 7.0, 8.0, and 9.0 is vulnerable to cross-site scripting. This vulnerability allows users to embed arbitrary JavaScript code in the Web UI thus altering the intended functionality potentially leading to credentials disclosure within a trusted session. IBM Reference #: 1997743
An issue was discovered in St. Jude Medical Merlin@home, versions prior to Version 8.2.2 (RF models: EX1150; Inductive models: EX1100; and Inductive models: EX1100 with MerlinOnDemand capability). The identities of the endpoints for the communication channel between the transmitter and St. Jude Medical's web site, Merlin.net, are not verified. This may allow a man-in-the-middle attacker to access or influence communications between the identified endpoints.
An issue was discovered in BINOM3 Universal Multifunctional Electric Power Quality Meter. Users do not have any option to change their own passwords.
An issue was discovered in Hanwha Techwin Smart Security Manager Versions 1.5 and prior. Multiple Path Traversal vulnerabilities have been identified. The flaws exist within the ActiveMQ Broker service that is installed as part of the product. By issuing specific HTTP requests, if a user visits a malicious page, an attacker can gain access to arbitrary files on the server. Smart Security Manager Versions 1.4 and prior to 1.31 are affected by these vulnerabilities. These vulnerabilities can allow for remote code execution.
An issue was discovered in Hanwha Techwin Smart Security Manager Versions 1.5 and prior. Multiple Cross Site Request Forgery vulnerabilities have been identified. The flaws exist within the Redis and Apache Felix Gogo servers that are installed as part of this product. By issuing specific HTTP Post requests, an attacker can gain system level access to a remote shell session. Smart Security Manager Versions 1.5 and prior are affected by these vulnerabilities. These vulnerabilities can allow for remote code execution.
An issue was discovered in certain legacy Eaton ePDUs -- the affected products are past end-of-life (EoL) and no longer supported: EAMxxx prior to June 30, 2015, EMAxxx prior to January 31, 2014, EAMAxx prior to January 31, 2014, EMAAxx prior to January 31, 2014, and ESWAxx prior to January 31, 2014. An unauthenticated attacker may be able to access configuration files with a specially crafted URL (Path Traversal).