Vulnerabilities
Vulnerable Software

Vulnerability Details CVE-2020-12037

Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.
Exploit prediction scoring system (EPSS) score
EPSS Score 0.001
EPSS Ranking 22.2%
CVSS Severity
CVSS v3 Score 7.5
CVSS v2 Score 5.0
Products affected by CVE-2020-12037


Contact Us

Shodan ® - All rights reserved