Vulnerability Details CVE-2017-12716
Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitoring units. Additionally, the Accent and Anthem pacemakers store the optional patient information without encryption. CVSS v3 base score: 3.1, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:N/A:N. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.
Exploit prediction scoring system (EPSS) score
EPSS Score 0.0
EPSS Ranking 10.4%
CVSS Severity
CVSS v3 Score 6.5
CVSS v2 Score 3.3
Products affected by CVE-2017-12716
-
cpe:2.3:h:abbott:accent:-
-
cpe:2.3:h:abbott:accent_mri:-
-
cpe:2.3:h:abbott:accent_st:-
-
cpe:2.3:h:abbott:anthem:-
-
cpe:2.3:o:abbott:accent_firmware:*
-
cpe:2.3:o:abbott:accent_mri_firmware:*
-
cpe:2.3:o:abbott:accent_st_firmware:*
-
cpe:2.3:o:abbott:anthem_firmware:*