Vulnerability Details CVE-2017-11578

It was discovered as a part of the research on IoT devices in the most recent firmware for Blipcare device that the device allows to connect to web management interface on a non-SSL connection using plain text HTTP protocol. The user uses the web management interface of the device to provide the user's Wi-Fi credentials so that the device can connect to it and have Internet access. This device acts as a Wireless Blood pressure monitor and is used to measure blood pressure levels of a person. This allows an attacker who is connected to the Blipcare's device wireless network to easily sniff these values using a MITM attack.
Exploit prediction scoring system (EPSS) score
EPSS Score 0.014
EPSS Ranking 79.7%
CVSS Severity
CVSS v3 Score 5.9
CVSS v2 Score 4.3