Vulnerability Details CVE-2015-3954
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Exploit prediction scoring system (EPSS) score
EPSS Score 0.005
EPSS Ranking 66.6%
CVSS Severity
CVSS v3 Score 9.8
CVSS v2 Score 10.0
Products affected by CVE-2015-3954
-
cpe:2.3:h:pifzer:plum_a+3_infusion_system:-
-
cpe:2.3:h:pifzer:plum_a+_infusion_system:-
-
cpe:2.3:h:pifzer:symbiq_infusion_system:-
-
cpe:2.3:o:pifzer:plum_a+3_infusion_system_firmware:13.6
-
cpe:2.3:o:pifzer:plum_a+_infusion_system_firmware:13.4
-
cpe:2.3:o:pifzer:symbiq_infusion_system_firmware:3.13