Vulnerability Details CVE-2015-3952
Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Exploit prediction scoring system (EPSS) score
EPSS Score 0.001
EPSS Ranking 31.7%
CVSS Severity
CVSS v3 Score 7.5
CVSS v2 Score 5.0
Products affected by CVE-2015-3952
-
cpe:2.3:h:pifzer:plum_a+3_infusion_system:-
-
cpe:2.3:h:pifzer:plum_a+_infusion_system:-
-
cpe:2.3:h:pifzer:symbiq_infusion_system:-
-
cpe:2.3:o:pifzer:plum_a+3_infusion_system_firmware:13.6
-
cpe:2.3:o:pifzer:plum_a+_infusion_system_firmware:13.4
-
cpe:2.3:o:pifzer:symbiq_infusion_system_firmware:3.13